- coronary angiography
- gastrointestinal studies (barium meal/swallow)
- intravenous pyelography
Building on the CT-120's strength as a cardiac imaging system, the fluoroscopy acquisitions can be cardiac or respiratory gated, or a combination of both.

40 - 125 kV
16 - 80 mA
6.3 - 200 ms
Full range of gating options:
ECG
RESP
ECG+RESP
For applications that require repeated imaging and may require imaging at multiple bed positions, the protocol definition allows you to specify an anode heat threshold and a table offset. The anode heat threshold defines the maximum heat level that is allowed before the acquisition will continue. If the current heat level is above this, the acquisition system will wait until it drops below the threshold (Note: heat level is continuously monitored and displayed - even during x-ray exposure). The table offset is a distance (mm) from the landmark position. Before the acquisition, the specimen cradle will be moved to this position. Both of these features streamline work flow and assist automation.

For more information, please contact us.